Laminated Glass Boil Test Device

Name

Boiling Test device for PVB interlayer

Function

Determining the effect of exposure of safety glazing materials to temperature and humidity condition by boiling.

Related Standard

  • EN ISO 12543 – 4 Glass in building -Laminated glass and laminated safety glass – Part 4: Test methods for durability – Clause 5 High-temperature test
  • AS/NZS 2208 Safety glazing materials in buildings – Appendix F Boil Test
  • ANSI Z97.1 Safety Glazing Materials Used in Buildings – Safety Performance Specifications and Methods of Test – Clause 5.3.2 – ( 1 ) Boil Test

Technical Data:

  • Outside Size: 500mm*300mm*580mm
  • Inner Size: 395mm*260mm*495mm
  • Working Voltage and Frequency : AC220V/50HZ
  • Power: 3KW
  • At most 6 samples with 300 mm

Apparatus

The apparatus include two water baths and one temperature control device, each water baths is of sufficient size to hold the test specimen fully immersed vertically on edge for the duration test.

Principle

 Specimens are immerse in hot water and boiling water for specified period of time and inspect bubbles and other defects

Working Environment:

  • Temperature: 20℃±5℃
  • Atmospheric Pressure: 8.60×104Pa-1.06×105Pa
  • Relative Humidity: 40%-80%

Instructions:

  • 1, Place the boil test equipment stably, and put the sample holder into the box;
  • 2, Pour water into box, control the water level is about 50-80mm from the top of the box, cover the box;
  • 3, Switch on the power supply, set the temperature-controlled meter to required temperature (63~69°C).
  • 4, When the temperature reach the required temperature (63~69°C), place the samples vertically on the sample holder inside the box,3 minutes later, take out the samples and quickly put them into another boiling test box, after 2 hours, take out the samples and observe.
  • 5, Disconnect the power supply, drain the water after cooling.

Attention:

  • 1,  check whether the box is deformed or damaged before use.
  • 2, check the water level in the box before switching on the power supply.
  • 3,  keep the box dry and maintain regularly when it is not used for a long time.
  • 4,  it is normal to have a little scale after heating for a long time. Please clean up in time so as not to affect the heating effect.

Packing details:

  • Plywood case (Meet import and export requirements)
  • Packing Size: 650mm*535mm*900mm
  • Total Weight: 49kg